Recall Z-2976-2024 ยท RDW General Purpose Laser Products

PACIFIC LASERTECH, LLC

25-LHP-928, HeNe Laser System, 25 mW

DateSep 10, 2024 (initiated Mar 1, 2024)
StatusOpen, Classified
ReasonThese laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
Root causeRadiation Control for Health and Safety Act
ActionThe firm notified all purchasers of the defect. Customers were instructed to return all units for replacement of the switch free of charge.
Quantity4
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.