Recall Z-2415-2011 ยท DYB Introducer, Catheter

Terumo Medical Corporation

Pinnacle Destination Guiding Sheath, 6 French, 90 cm, Straight Tip, Tuohy-Borst Valve, Lot MD24, MD31, ME07, & ME14. 510(k) K012812 Guiding Sheath is used to facilitate the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal arteries for the Renal Guiding Sheath and the carotid arteries for the Carotid Guiding Sheath.

DateJun 6, 2011 (initiated Jun 24, 2010)
StatusTerminated
ReasonMolding defect in destination product that render the device unusable.
Root causeEquipment maintenance
ActionTerumo Medical Corporation sent an Urgent Product Bulletin Recall letter dated June 24, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Terumo requested customers to immediately discontinue use, and quarantine any available stock of the listed product codes. In order to facilitate the exchange of these devices, a Terumo Territory Manager contacted all customers by July 9th to confirm receipt of the Terumo Destination Bulletin and scheduled an appointment to visit each customer in person by July 31st to personally
Quantity214 units
DistributionWorldwide Distribution - USA including AL, AK, AZ, TX, LA, MS, MN, MO, CA, GA, FL, HI, KS, IA, ID, IL, TN, NM, OR, OK, WA, and WI and the countries of Belgium, Brazil, and Panama
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012812CAROTID GUIDING SHEATH
Terumo Medical Corp., Nov 14, 2001