Recall Z-2383-2011 ยท KNY Accessories, Catheter, G-U

Gyrus Acmi, Incorporated

Gyrus ACMI, REF 25 BX, PTFE Coated 3 cm Flexible Tip Guidewire, .035" x 150 cm, Sterile EO, Rx Only, Gyrus ACMI, Inc. 136 Turnpike Road,Southborough, MA 01772-2104 USA Ureteral guidewires are used to establish access in the urinary system for the passage of urological devices. An introducer is first used to establish entry to various portions of the urinary tract (urethra, bladder, ureter), the

DateMay 26, 2011 (initiated Apr 15, 2011)
StatusTerminated
ReasonGyrus ACMI, Inc. wants to bring a recall issue to their customers attention concerning one lot of their 25BX-PTFE Coated 3 cm Flexible Tip Guidewire. As a result of a manufacturing error in the production this lot of guidewires, some units of product were loaded in the dispenser backwards. In such cases, the stiff end of the guidewire would be dispensed from the product first and, if undetected, could be inserted into the patient. Depending on the specific clinical circumstances of the patient, this could potentially cause perforation of the ureter and surrounding anatomy. Therefore Gyrus
Root causeProcess control
ActionThe firm, Olympus ACMI, sent an "Urgent: Medical Device Recall" letter dated April 15, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1) Inspect their stock to confirm if any of the affected lot of products remains in their possession 2)Immediately cease any further use of the affected product, remove, quarantine and return it. 3) Call their Gyrus ACMI customer service representative at 1-888-542-7266 to obtain a Returned Goods Authorization 4) Return the product to: Attn: PTFE Guidewire Recall RG
Quantity145
DistributionNationwide distribution: USA including states of: CA, GA, HI, IN, KY, MT, NY, NC, PA, TX, UT, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.