Recall Z-2403-2011 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Capsule Tech Inc.

Capsule Neuron UMPC is a component in the DataCaptor" Connectivity System. Catalog Number: DC-NU-MPC Capsule Neuron" UMPC is a bedside device that enables automatic collection of vital signs data that can local cache data and connect to the hospital network through a wired Ethernet connection or via a standard 802.11 b/g/n wireless network, available in a fixed or mobile mounting option, running

DateJun 1, 2011 (initiated Aug 25, 2010)
StatusTerminated
ReasonSoftware: Data cached on the Neuron due to server unavailability is not being delivered to the server after server availability is restored. If a compact flash failure occurs, the file system may be corrupted and result in system hangs, lock-ups or crashes. Batteries can discharge completely in a Capsule Neuron" UMPC even though it is connected to power source
Root causeSoftware design
ActionCapsule Tech initiated three corrective action notifications for the Neuron" UMPC units three failure modes as follows: a) Data cached on the Capsule Neuron" UMPC due to server unavailability is not being delivered to the server after server availability is restored: i) A customer advisory was provided to a subset of consignees, along with DataCaptor software version 6.6.5 on December 1, 2010. ii) A customer advisory was provided to the remaining consignees, along with DataCaptor software version 6.6.7 on January 27, 2011. If a compact flash failure occurs, the file system may be corrupted and
Quantity3,144 Capsule Neuron UMPC units
DistributionWorldwide Distribution - USA (nationwide) and the country of Canada