Recall Z-2181-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19;

DateMay 10, 2011 (initiated Apr 5, 2011)
StatusTerminated
ReasonPosition sensor failure mode may result in an inaccurate position calculation. 1. The video returned by one of the two cameras inside the Spectra is all white or all black, and the Spectra stops tracking. Or 2. Certain video intensities are not available on one of the two cameras. The image has abnormally abrupt transitions from dark to light portions, without the normal shades of gray.
Root causeNonconforming Material/Component
ActionVarian Urgent Medical Device Correction, Urgent Field Safety Notice letters were sent to all affected facilities by certified mail on April 5, 2011.
Quantity20
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092871TRILOGY MX
Varian Medical Systems, Inc., Nov 30, 2009