Recall Z-2222-2011 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Colleague, Colleague CX and Colleague CXE Infusion Pumps, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division. Product Codes: 2M8151, 2M8161 and 2M9161. Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of soluti

DateMay 19, 2011 (initiated Apr 14, 2011)
StatusTerminated
ReasonThe Hi-Pot safety testing may not have been conducted on the affected pumps during service, prior to being returned to the end user.
Root causeEmployee error
ActionThe firm, Baxter Healthcare Corporation, telephoned the customers on April 14, 2011, informing them that that the Hi-Pot test,and sent a follow up "URGENT DEVICE CORRECTION" letter dated April 27, 2011, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to locate the pumps and take them out of service as soon as possible (Baxter will then send a shipping container for retrieval, re-inspection and return of the device back to the customer at no cost.); and return their colleague pumps for full functional testing including the HiP
Quantity117 pumps
DistributionWorldwide distribution: USA (nationwide) including states of :Alabama, Arizona, California, Connecticut, Florida, Georgia, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Mississippi, North Carolina, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Texas, Virginia, Washington and We
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063696COLLEAGUE VOLUMETRIC INFUSION PUMP MODEL# 2M8151,8153,8161,8163,9161,9163
Baxterhealthcare, Inc., Feb 27, 2007