Recall Z-2241-2011 ยท JAK System, X-Ray, Tomography, Computed

Philips And Neusoft Medical Systems Co., Ltd.

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part: "***Neusoft***NeuViz16 MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED FOR NEUSOFT MEDICAL SYSTEMS CO., LTD***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD***ADDRESS: NEUSOFT PARK, HUNNAN INDUSTRIAL AREA, SHENYANG***CHINA***".

DateMay 23, 2011 (initiated Feb 9, 2011)
StatusTerminated
ReasonDimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image.
Root causeOther
ActionNeusoft Medical Systems sent an Urgent Device Correction Notification letter dated February 9, 2011, via certified mail to their distributor NEUISYS. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to discontinue use of combined images. Instead, use MPR Batch for diagnostic reviews until the planned upgrade is provided to the product. Neusoft Medical System and Philips Medical System released the mandatory Field Change Order (FCO) by March 31, 2011. All nine affected systems will be updated with aforementioned kits in six
Quantity9 units distributed in US
DistributionNationwide Distribution including CA, GA, LA, NC, OH, TN and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090173NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Feb 5, 2009
K092742NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Sep 22, 2009