Recall Z-2692-2011 ยท LLZ System, Image Processing, Radiological

Eos Imaging

Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850. EOS system is a digital radiography system used in general radiographic examinations. sterEOS workstation is a med

DateJun 28, 2011 (initiated Apr 13, 2011)
StatusTerminated
ReasonPrint true size, print requests from the EOS acquisition station or the sterEOS image review station may, in certain cases, be affected by a different enlargement factor during printing. T
Root causeSoftware design
ActionEOS imaging sent a "FIELD SAFETY NOTICE" letter dated April 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to wait for the printing queue to empty before sending a second "print true queue" request with different print settings. Customers should then click the button shown on the instruction sheet to view the print queue. EOS imaging will provide a software update to correct all affected systems at no cost to the customers. An EOS representative will contact customers to schedule a
Quantity5 EOS Systems, 10 sterEOS workstations
DistributionWorldwide Distribution--USA (nationwide) including the states of CA, IA, MO, VA, and WI, and countries of Canada and Europe.