ClearanceBrief
Recall Z-2737-2011 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Large Roller Pump 6" Diameter, Catalog 816571, Terumo Cardiovascular Systems, Corp., Ann Arbor, MI The product is indicated for up to 6 hours of extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures when used by a qualified medical professional.

DateJul 6, 2011 (initiated Mar 31, 2011)
StatusTerminated
ReasonThe firm has received multiple reports of the lower roller pump tube clamp mechanism failing to function in a small population of devices, making it difficult to remove or insert the tubing. If the failure occurs, the clamp dial will no longer be able to control opening/closing of the tube clamp mechanism.
Root causeDevice Design
ActionTerumo Cardiovascular Systems sent an URGENT MEDICAL DEVICE CORRECTION letter to all affected customers dated March 21, 2011. The letter identified the product, the problem and the action needed to be taken by the customer. An additional URGENT MEDICAL DEVICE RECALL correction notification and cover letter was sent to all consignees on XX/XX/XX to explain the reason for the expansion of the recall and will provide an update to the timing of implementation of the field correction. Expansion now includes devices manufactured between 2006 and March 2012. The correction strategy is to issue th
Quantity1988 devices.
DistributionWorldwide Distribution -- USA (nationwide) and the countries of Argentina, Australia, Belgium, Canada, Chile, Columbia, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Jordan, Korea, Kuwait, Malaysia, Mexico, Pakistan, Republic of Georgia, Russia, Sa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003