ClearanceBrief
Recall Z-2713-2011 ยท LXH Orthopedic Manual Surgical Instrument

Biomet, Inc.

Comprehensive Shoulder Instruments Fracture Stem Inserter w/Version Control, REF 31-406901, Biomet Orthopedics Warsaw, IN, Instrument used to introduce the stem into the canal. The instrument is used to insert the shoulder implant into the shoulder.

DateJun 30, 2011 (initiated Apr 7, 2011)
StatusTerminated
ReasonWhen the surgeon attempts to remove the inserter from the stem, the inserter may become stuck not allowing it to disengage from the stem. The stem may have to be removed from the patient in order to disengage the inserter from the stem. If the stem is removed from the patient, it could potentially cause a significant delay in the procedure or damage/fracture of the humerus which may require inter
Root causeNonconforming Material/Component
ActionThe firm, Biomet Orthopedics, sent two "URGENT MEDICAL DEVICE RECALL NOTICE" letters one dated April 7, 2011 and the second letter dated April 12, 2011 which included an additional lot number, to its customers. The letters described the product, problem and actions to be taken. The customers were instructed to immediately locate and discontinue use of the noted product; carefully follow the instructions on the enclosed "FAX Back Response Form"; complete and return the Fax Response Form within three (3) business days via fax to 574-372-1683; use priority carrier for their shipment; If the pr
Quantity469 both products
DistributionWorldwide distribution: USA (nationwide) including Puerto Rico; and countries of: Australia, Canada, Chile, Columbia, Costa Rica, Italy, Japan, Korea, Netherlands, New Zealand, Spain and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.