Recall Z-2243-2011 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

ARTISTE , ONCOR and PRIMUS digital accelerators with syngo RT Therapist v4.1 and 4.2 system; Therapist Express Basic, syngo RT Therapist and syngo RT Therapist Connect/MOSAIQ OIS. The ONCOR and PRIMUS may be updated to utilize the RTT v4.x software. Distributed by Siemens, Healthcare, Concord, CA 94520 Radiation therapy

DateMay 23, 2011 (initiated Mar 31, 2011)
StatusTerminated
ReasonSiemens found a potential safety issue related to patient treatment with table offsets on syngo RT Therapist (RTT) 4.1 and 4.2 which could result in dose being delivered to wrong location.
Root causeSoftware design
ActionSiemens initiated an Urgent Medical Device Correction letter on March 31, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: use only relataive table setup with all planned table values '0' if applicable. In cases where the treatment plan contains at least one table parameter other than '0' to consider the following: Use only one PVG plus resulting offset per fraction. Once an offset has been applied do not use 'apply offset' at RTT for any further offset. Select the PVG and treatment bea
Quantity164 active devices
DistributionWorldwide Distribution - USA including CA, DE, FL, KY, LA, MA, MI, MO, NC, NE, NJ, NY, OH, PA, SD, TN, UT, WI, and WV and the countries of Australia, Belgium, Canada, Croatia, Egypt, France, Germany, Hungary, India, Ireland, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Phili
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006
K072486MODIFICATION TO HEMCON BANDAGE AND HEMCON BANDAGE OTC
Hem Con Medical Technologies, Inc., Aug 6, 2008
K993425PRIMUS
Siemens Medical Solutions USA, Inc., Nov 26, 1999