Recall Z-2190-2011 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

ARTISTE / RT Therapist v4.1 system. The ONCOR and PRIMUS system may be updated to utilize the RT Therapist v4.1 software. Siemens Healthcare, Concord, CA. The intended use is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

DateMay 12, 2011 (initiated Mar 24, 2011)
StatusTerminated
ReasonWhen using the ARTISTE Linac and RT Therapist v4.1 along with an imported cropped reference image, the image maybe processed incorrectly, thus resulting in a mistreatment.
Root causePending
ActionDistribution of the Urgent: Medical Device Correction Customer Advisory Notice began on March 24, 2011 via certified mail or by the Siemens Customer Service Group. The letter identified the issue and when it occurs. Users were asked to verify the isocenter position of the reference image in the Portal Review application before using a reference image imported from the treatment planning system. Siemens is currently working on a solution and will notify customers as soon as the update is available. The letter should be included with the Owner Manual of the system in the chapter for "Safety
Quantity228 active devices
DistributionWorldwide Distribution
FDA recordOpen on fda.gov