Recall Z-2205-2011 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance iCT, Computed Tomography X-Ray System, including the Brilliance iCT SP system, Model Numbers: 728311 and 728306. The affected systems are equipped with the following software version: 3.2.0.19010. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

DateMay 19, 2011 (initiated Mar 7, 2011)
StatusTerminated
ReasonA program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.
Root causeOther
ActionA 'Field Safety Notice - Urgent - Medical Device Correction' notification was sent to the firm's customers via Certified mail on 3/8/2011. The notification letter explains the nature of the Metal Artifact Reduction/Metal Artifact Correction software error; the Brilliance CT X-Ray systems subject to correction and how to identify them; and the circumstances under which this error can occur. It also asks that the user stop running the MAR application on any of the affected Brilliance systems identified with the recalled serial numbers and software versions alluded to in the medical device cor
Quantity8 units
DistributionNationwide Distribution -- DC, IL, IN, MI, MO, NY, OH, OR, PA, TX, and VT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060937BRILLIANCE VOLUME
Philips Medical Systems (Cleveland), Inc., Jun 5, 2006