Recall Z-2150-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

Siemens brand ACUSON S2000, Model 10434730, ABVS ultrasound systems; Affected sub-components Numbers: (1) 10434968, Boom Welding (for S/Ns 250157, 250162, 250163, 250164, 250166, 250167 and 250169); (2) 10434969, Arm Welding (for S/N's 250171, 250173, 250177); Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Concord, CA The S2000 Automated Breast Volume Scanner (

DateMay 6, 2011 (initiated Mar 28, 2011)
StatusTerminated
ReasonThe weld connecting the ABVS arm to the system base has the potential to fail, causing the arm to fall off of the unit.
Root causeProcess change control
ActionSiemens Medical Solutions USA, Inc. initiated the recall via phone calls and e-mail on March 28, 2011, to all affected customers, informing them of the affected product and providing instructions on the recall. Three units S/N 250171, 250173, and 250177 were to be removed and shipped back to Siemens. Siemens Service personnel were to be dispatched to the customer sites to perform validated stress test on units S/N 250157, 250162, 250163, 250164, 250166, 250167, and 250169 at customer sites. For questions regarding this recall call 650-694-5993.
Quantity10
DistributionWorldwide Distribution - USA, Portugal, Germany, Russia, Estonia, Brazil, France, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081148SIEMENS ACUSON 2000 ULTRASOUND SYSTEM WITH ABVS TRANSDUCER
Siemens Medical Solutions USA, Inc., Ultrasound DI, Dec 22, 2008