ClearanceBrief
Recall Z-2752-2011 ยท GKZ Counter, Differential Cell

Abbott Laboratories

Abbott CELL-DYN 1700 Analyzer, CELL-DYN 1700 CS (Closed Sample).Manufactured by Abbott Diagnostics Division, Santa Clara, Ca Medical Device, hematology analyzer for in vitro diagnostic use in clinical Laboratories.

DateMar 17, 2011 (initiated Mar 27, 2009)
StatusTerminated
ReasonExpanded Recall to include these additional products: FIRE HAZARD-- Some power supply modules may have incorrect fuses installed, which may result in fires and smoking.
Root causeComponent design/selection
ActionAbbott Laboratories sent Product Correction letters, dated March 22, 2010, to its customers. The letter listed the additional products that are now included in the recall and gave an explanation of the reason for recall. The letter also discussed the safety impact and provided a list of necessary actions for customers. Customers are to check that the ampere rating of the fuse matches the voltage of operation that is currently being utilized with their CELL-DYN 3000 Analyzer. If the fuse is correct, then the steps provided should be followed. If the fuse is incorrect, then customers are to
Quantity2019 CELL-DYN 1700 and 1700CS distributed worldwide, 896 of those in the US.
DistributionWorldwide Distribution -- USA, including states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of Argentina, Australia, Bahamas, Barbados, Brazil
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.