Recall Z-3162-2011 ยท NKB Thoracolumbosacral Pedicle Screw System

Orthofix, Inc.

Ratcheting Torque Limiting Handle, Catalog Number 52-1012, Orthofix Spinal Implants, Firebird Spinal Fixation System, Firebird Spinal Fixation System DDD Instrument Case. Product is intended for use with the Set Screw Driver (52-1061) and Adjustable Counter Torque Wrench (52-1265) to perform the final tightening of set screws in a spinal implant construct.

DateSep 7, 2011 (initiated Apr 1, 2011)
StatusTerminated
ReasonUpon reaching torque limit, there is a possibility the product may jam or exhibit a ratchet direction switch from forward to reverse.
Root causeDevice Design
ActionOrthofix, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 1, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify and remove the affected product from use. Customers were advised to use the combination of the Ratching R Handle and the Set Screw Driver if they experience problems with the Ratching Torque Limiting Handle. A Tracking and Verification form was attached for customers to complete and return to the Orthofix Inc. Regulatory Affairs Department via fax a
Quantity287 units
DistributionWorldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092624FIREBIRD SPINAL FIXATION SYSTEM
Blackstone Medical, Inc., Sep 25, 2009