Recall Z-2718-2011 ยท HXI Passer, Wire, Orthopedic
Interventional Spine Inc
PERPOS PLS SYSTEM, CATALOG NO. 9045-01 A sterile, medical device kit consisting of the components listed in Table 1-1 below. The kit is used by medical professionals to stabilize the spine as an aid to fusion through immobilization of the facet joints. A Surgical Technique brochure is provided in Appendix A and other labeling, including instructions for use, is provided in Appendix B.
| Date | Jul 1, 2011 (initiated Mar 23, 2011) |
| Status | Terminated |
| Reason | This recall has been initiated due to K-wires with a distal tip diameter that may be greater than specified causing an interference fit with the canulated screw. Removing the K-wire during a procedure may dislocate the implant resulting in the implant loosening or being pulled out or position. Use of this product may result in a delay of the surgical procedure and/or pull-out of the facet screw |
| Root cause | Pending |
| Action | Interventional Spine sent an RECALL NOTIFICATION letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. The customer was instructed to take the following actions: Immediately examine inventory and quarantine product subject to recall. In addition, if this product was further distributed, please notify the Vice President Quality Assurance at once. Consignee were asked to also complete and return the enclosed response form as soon as possible. Consignee with questions were directed to the Cons |
| Quantity | 864 units total for all products |
| Distribution | Nationwide Distribution including the states of to NC, NY, TN & VA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.