Recall Z-0045-2012 ยท HXX Screwdriver

Interventional Spine Inc

PERPOS PLS System, Single Use; CATALOG #'s 9045-01 & 9045-02. Interventional Spine, Inc. A sterile kit used to stabilize the spine as an aid to fusion through immobilization of the facet joints.

DateOct 13, 2011 (initiated Feb 3, 2011)
StatusTerminated
ReasonThis recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure
Root causeNonconforming Material/Component
ActionInterventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006.
Quantity2,620 units
DistributionWord wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.