Recall Z-0045-2012 ยท HXX Screwdriver
Interventional Spine Inc
PERPOS PLS System, Single Use; CATALOG #'s 9045-01 & 9045-02. Interventional Spine, Inc. A sterile kit used to stabilize the spine as an aid to fusion through immobilization of the facet joints.
| Date | Oct 13, 2011 (initiated Feb 3, 2011) |
| Status | Terminated |
| Reason | This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure |
| Root cause | Nonconforming Material/Component |
| Action | Interventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006. |
| Quantity | 2,620 units |
| Distribution | Word wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K. |