Recall Z-2125-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

ARIA, RT Chart Application v 8.0 through 8.9 Varian Medical Systems, Palo Alto, CA. Aria Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment.

DateMay 2, 2011 (initiated Mar 9, 2011)
StatusTerminated
ReasonAn issue was identified with the gantry rotation direction when creating opposing field within the RT Chart function of ARIA, or possibly with third party planning systems and delivering the plan at 4DITC.
Root causeSoftware design
ActionVarian Medical Systems, Inc. Oncology Systems sent Urgent Medical Device Correction Field Safety Notice letters on March 9, 2011, to all affected customers. ***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letters identifeid the product, the problem, and the action to be taken by the customer. Customers were instructed to: Manually enter the gantry angel 0.0 when the plan indicates angel 0.0. Manual entry is only allowed to one decimal place, so the range that triggers the issue cann
Quantity612 units
DistributionWorldwide Distribution - USA including CA, DC, FL, GA, MD, MN, MS, NY, OH, and TN and the countries of Canada, Australia, Austria, Bahrain, Bagladesh, Belarus, Belgium, Brazil, Columbia, Costa Rica, Cypress, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Hon
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093527ARIA RADIATION ONCOLOGY WITH MIRS
Varian Medical Systems, Inc., Dec 15, 2009