ClearanceBrief
Recall Z-2236-2011 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Oridion Medical 1987 Ltd.

Micro Stream Filterline Infant/Neonatal products: Part # 010807 - VlTALINE H SET INFANT/NEONATAL 25UN (Inf/Neo Airway Adapter). The adult/pediatric airway adapter and the adult/pediatric FilterLine and VitaLine sets are not affected by this recall. The intended use of Microstream Filter Lines is to convey a sample of a patient's expired breath to a capnograph in order to measure the CO2 partial

DateMay 25, 2011 (initiated Mar 25, 2011)
StatusTerminated
Reasonfine plastic strands on inner surface of the infant neonatal airway adapter may become dislodged and inhaled by the patient
Root causeEquipment maintenance
ActionThe firm, Oridion, sent an "Microstream Filterline Field Correction Notice" letter dated March 25, 2011 to its US and foreign customers via telephone, email with the documentation discussed in the call, and letter. The letter described the product, problem and actions to be taken. The customers were instructed to place all affected product held in their inventory on ship hold; gather these units in their inventory and at the customer sites and exchange with the new product; inspect filter line starter (sample) kits shipped with Capnostream 20 and Microcap patient monitors for affected produc
Quantity300 units in US (12 boxes with 25 units ea)
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.