Recall Z-2419-2011 ยท DQO Catheter, Intravascular, Diagnostic

Boston Scientific Corporation

Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY 2011-05, REF 39014, UPN H749390140, Sterile R Sterilized using irradiation, Made in USA 47900 Bayside Parkway, Fremont, CA 94538 The intended use as stated in Japan Shonin 21200BZY00484000: This product is an imaging catheter with built-in ultrasonic transducer that sends ultrasound waves inside

DateJun 10, 2011 (initiated Mar 28, 2011)
StatusTerminated
ReasonCatheter tip detachment that is not associated with elongation of the catheter typically caused by excessive pulling forces. EXPANDED: Boston Scientific is initiating a recall of all iCross Coronary Imaging Catheters because they have received and confirmed 8 occurrences of catheter tip detachments due to embrittlement.
Root causeProcess design
ActionBoston Scientific " Important Customer Notification" letters dated March 14, 2011 were hand delivered to customers by Boston Scientific sales representatives starting on March 28th, 2011. The sales representatives verbally reviewed information. A tracking log will be maintained document completion of the customer notifications. The letter described the product, problem and possible root cause as well as one corrective action, a change in sterilization method from Gamma to E-beam implementation. Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter
Quantity30,669
DistributionWorldwide Distribution - USA (nationwide) and Puerto Rico and the countries of Japan, Trinidad, Tobago, and American Virgin Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050577ATLANTIS SR PRO2 CORONARY IMAGING CATHETER, MODEL 39014
Boston Scientific Corp, Mar 30, 2005
K063312MODIFICATION TO ATLANTIS SR PRO 2 AND SR PRO CORONARY IMAGING CATHETERS, MODEL 39014 AND 38942
Boston Scientific Corp, Nov 30, 2006