Recall Z-2048-2011 ยท DYB Introducer, Catheter

Arrow International Inc

Percutaneous Sheath Introducer Kit Intended usage: Permits venous access and catheter introduction to the central circulation system

DateMar 25, 2011
StatusTerminated
ReasonLabeling discrepancy - the outer corrugate label as well as a card inside the package state the product contains dry natural rubber - this is not duplicated on the unit packaging.
Root causeOther
ActionThe firm, Arrow International, sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated March 25, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately quarantine any products with the listed lot numbers; complete and return the Immediate Attention Request form via fax to 800-343-2935 indicating how much product they have in stock and whether they need labels. In the event that an alternative approach is needed, the customers are to contact Arrow Customer Service for more information. Once the labels have been
Quantity840
DistributionNationwide distribution: USA including states of: PA and NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K780532PERCUTANEOUS SHEATH INTRODUCER KIT
Arrow Intl., Inc., Apr 21, 1978