Recall Z-2053-2011 ยท MRA Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented

Smith & Nephew Inc

R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 32 MM I.D. 48 MM O.D., REF 71338948, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06 Usage: Primary total hip arthroplasty due to non-inflammatory arthritis (degenerative joint disease) such as osteoarthritis, avascular necrosis or traumatic arthritis

DateApr 22, 2011 (initiated Mar 11, 2011)
StatusTerminated
ReasonDuring the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.
Root causeEmployee error
ActionAll affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall. Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.
Quantity296 units
DistributionAK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.