Recall Z-2128-2011 ยท GKZ Counter, Differential Cell
Beckman Coulter Inc.
UniCel DxH 800 Coulter Cellular Analysis System Part Number: 629029 The UniCel DxH 800 Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel DxH 800 Analyzer provides: - Complete Blood Count (CBC), Leukocyte 5 Part Differential (Diff), Reticulocyte (Retic) and Nucleated Red Blood Cell (NRBC
| Date | May 2, 2011 (initiated Oct 20, 2010) |
| Status | Terminated |
| Reason | Body Fluids results may be incorrectly interpreted by the Laboratory Information System (LIS). Body Fluid results ~ 1,000 celis/uL are reported using the US numeric separators to demark thousandths places (comma, ",") and partial units (decimal, "."). An LIS that uses non-US formats could interpret the comma as a partial units separator, which could lead to reporting erroneously low body fluid r |
| Root cause | Other |
| Action | The firm, Beckman Coulter, issued an "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated October 20, 2010, with PCA Response Form to all customers who purchased the UniCel DxH 800 Coulter Cellular Analysis System. This letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1) Remove all cassettes from the mix station prior to initiating the Diluent Dispense function. Do not initiate Diluent Dispense while specimens are being processed in cassette presentation. 2)Do not use pre-diluted samples for RETIC counts. The instrument is v |
| Quantity | 422 units in total (175 in US) |
| Distribution | Worldwide distribution: USA (nationwide) and countries including:Australia, Belgium, Canada, Colombia, Croatia, France, Germany, Greece, Hong Kong, India, Italy, Japan, Korea, Republic of, Macao, Malaysia, Myanmar, Netherlands, Norway, Philippines, Puerto Rico, Qatar, Russian Federation, Saudi Arabi |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K081930 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM Beckman Coulter, Inc., Dec 19, 2008 |