Recall Z-0158-2015 ยท DTR Heat-Exchanger, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Cardioplegia Delivery Set- Item Number: 16130 Product Usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery.

DateOct 31, 2014 (initiated Mar 1, 2011)
StatusTerminated
ReasonDefect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line
Root causeDevice Design
ActionTerumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to rep
Quantity777 units
DistributionWorldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.