ClearanceBrief
Recall Z-2024-2011 ยท KDC Instrument, Surgical, Disposable

Hospira Inc.

Hospira Latex-Free Blood Secondary Set, 26 Inch with Secure Lock; a sterile Rx tubing set for the administration of fluids from a blood bag in series with an I.V. solution bottle and I.V. set; Made in Costa Rica; Hospira, Inc., Lake Forest, IL 60045 USA; List No. 4602-58 For the administration of fluids from a blood bag in series with an I.V. solution bottle and I.V. set

DateApr 19, 2011 (initiated Mar 28, 2011)
StatusTerminated
ReasonThere is a potential for the air filter at the distal end of the set to become detached. This could result in the contamination of the sterile fluid path inside the tubing.
Root causeOther
ActionThe firm, Hospira, sent an "URGENT DEVICE RECALL" letter dated March 28, 2011 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to check their inventory and immediately quarantine any affected product; complete and return the Reply Form via fax to 1-866-324-3734, even if you do not have the affected product; inform healthcare professionals in their organization of this recall; if they have distributed the product further, notify their accounts and have them fax the reply form to 1-866-324-3724; return the affected product to St
Quantity28,608 sets
DistributionWorldwide distribution: USA (nationwide) and countries including: Canada, Korea and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K780880Y-TYPE BLOOD SET
Abbott Laboratories, Jun 30, 1978