Recall Z-1930-2011 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

GE Healthcare, CARESCAPE Monitor B650 The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B650 is indicated for monitoring and recording of

DateApr 8, 2011 (initiated Mar 23, 2011)
StatusTerminated
ReasonGE Healthcare has become aware of a potential safety issue associated with the CARESCAPE Monitor B650 & B850 Cardiac Output measurement using Patient Data Module (PDM), and activation of ECG Leads Off alarm in specific cases. 1. When the CARESCAPE Monitor is used to measure Cardiac Output values with the Patient Data Module (PDM) or TRAM Module, Cardiac Output values may be calculated incorrectl
Root causeSoftware design
ActionThe firm, GE Healthcare, sent "Urgent Medical Device Correction" letters dated March 23, 2011 to its consignees/customers. The letters were addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, and Director of Biomedical Engineering. The letters described the Safety Issues, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. The customers were instructed to follow the safety instructions documented in the letter. GE Healthcare is working on a software correction that will be provided to their customers once it is available and will c
Quantity1443
DistributionAL, AZ, AR, CA, CO, CT, FL,GA, ID, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA,PR, SC, TN,TX, UT, VA, WA, WV, WI. UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, THAILAND, SWITZERLAND, SWEDEN SLOVENIA, SLOVAKIA, SPAIN, SINGAPORE, SAUDI ARABIA, QATAR
FDA recordOpen on fda.gov