Recall Z-2593-2011 ยท MUJ System, Planning, Radiation Therapy Treatment

Computerized Medical Systems Inc

XiO Radiation Treatment Planning System, XiO Release 2.1.0 and above Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.

DateJun 20, 2011 (initiated May 27, 2010)
StatusTerminated
ReasonXiO Software: The Patient Scan position is being confused with the treatment orientation of the patient. The Patient Scan position is determined from the data entered in PFM. Subsequently, when a New Teletherapy Plan is created for a patient in Teletherapy, the user is prompted with Head/Foot Gantry on the text page. This prompt is asking for the orientation of the patient when they are treated.
Root causeSoftware design
ActionComputerized Medical Systems Inc sent an undated Safety Notice to all affected customers. The notice indicated the problem, the clinical impact, and the workaround. The notice stated the sites will be notified when a solution is available.
Quantity1400
DistributionWorldwide Distribution - USA (Nationwide distribution, including Washington, DC and Puerto Rico) Foreign distribution was made to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Ri
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092132XIO RTP SYSTEM
Computerized Medical Systems, Inc., Sep 24, 2009