Recall Z-2596-2011 ยท MUJ System, Planning, Radiation Therapy Treatment

Computerized Medical Systems Inc

Focal Radiation Treatment Planning System, Focal Release 4.3.0 through 4.34.01 For use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the R

DateJun 20, 2011 (initiated Nov 30, 2007)
StatusTerminated
ReasonFocal: when a plan is sent to Focal from XiO and the structure properties are edited in Focal CT Sim activity, these changes result in changes to the MLC or field size for the plan when they should not. XiO will recalculate dose when modified plan is returned to XiO however the user may not notice that it has changed form the plant they had approved earlier. Editing the structure properties (col
Root causeSoftware design
ActionComputerized Medical Systems Inc sent a Customer Advisory, dated 11/30/07, was sent to consignees December 2007 to notify them of the issue and how to avoid the workaround. The advisory listed the problem, clinical impact, and a workaround. The firm will be sending each site a reminder that the issues is resolved.
Quantity233
DistributionWorldwide Distribution - USA (Nationwide AL, AR, CA, DE, FL, GA, IL, IN, MA, MD, MI, MOP, NE, NY, OH, PA, RI, TX, VA, WA, WI, and WV and the countries of Albania, Argentina, Belgium, Bosnia and Herzegovina, Brazil, China, Costa Rica, Croatia, Czech Republic, France, Germany, Greece, India, Italy, Ja
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013112FOCAL SIM
Computerized Medical Systems, Inc., Dec 5, 2001