Recall Z-2385-2011 ยท JEY Plate, Bone

Biomet Microfixation, Inc.

***BIOMET MICROFIXATION BIOMET MICROFIXATION 1520 TRADEPORT DRIVE JACKSONVILLE, FL 32218 USA www.biometmicrofixation.com***6 HOLE REGULAR DOUBLE Y PLATE***"1.5MM SYSTEM PLATE DOUBLE Y-SHAPE, REGULAR 0.6 MM TITANIUM***NON-STERILE PRODUCT*** This titanium plate is used in the 1.5mm Lorenz Midface System which is indicated for use in trauma, neurosurgery, craniomaxillary reconstruction and orthognat

DateMay 27, 2011 (initiated Jul 28, 2010)
StatusTerminated
ReasonBiomet Microfixation, Jacksonville, FL is recalling their 1.5 6 Hole Reg Double Y, Lots 785710, 753120, 693500 and 733600. This product is labeled as 01-7110-K (1.5 6 Hole Reg Double Y), however the package contains a 01-7112-K (1.5.6 Hold Long Double Y).
Root causeLabeling mix-ups
ActionThe firm, Biomet Microfixation, sent an "URGENT MEDICAL SAFETY ALERT NOTICE" letter dated July 28, 2010 to its customers/distributors. The letter described the product, problem and actions to be taken. The customers were instructed to inspect their stock and ensure that the plates were properly distributed into their trays and that they have the sufficient quantity of plates that they need. In addition, they were instructed to complete and return the attached Inventory Reconciliation form via fax to: Attn: Customer Relations Specialist at 904-741-9425 and notify the recipient of this notice, i
Quantity180
DistributionWorldwide distribution: USA (nationwide) including states of: FL, IN, KY, MT, NC, NE, NY, OH, TX, VA, and WI; and countries including: Argentina, Canada, Colombia, Czech Republic, Japan, Korea, Mexico, Norway, South Africa, Spain and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953385LORENZ 1.0MM, 1.5MM, 2.0MM SYSTEM
Walter Lorenz Surgical, Inc., Sep 28, 1995