Recall Z-2051-2011 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound, 1230 Shorebird Way Intended for the following applications: Fetal, abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal and Peripheral Vascular Applications.

DateApr 22, 2011 (initiated Mar 15, 2011)
StatusTerminated
ReasonThe Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report.
Root causeSoftware design
ActionSiemens issued a Customer Safety Advisory Notification letter to all customers on March 15, 2011, regarding a software error on the ACUSON S2000 ultrasound systems affected measurement . They are in the process of releasing a software update that will correct this issue. Siemens recommends reviewing the patient report at the end of each study to confirm that the report contains only measurements relevant to the exam just performed. No patient injury has been reported. Further questions regarding this recall please call (650) 694-5398.
Quantity1533 units
DistributionWorldwide Distribution including USA, states of AZ, VA, CA, NE, IL, MN, TN, FL, NM, and DC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082142ACUSON S2000 DIAGNOSTIC ULTRASOUND SYSTEM
Siemens Medical Solutions USA, Inc., Ultrasound DI, Nov 13, 2008