Recall Z-2008-2011 ยท GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
Terumo Cardiovascular Systems Corporation
Sarns Flexible Weighted Sucker in PUMP & TABLE PACK with PALL (custom pack), catalog number 70579-05, Sterile and Non-pyrogenic. Sterilized by ethylene oxide. Terumo Cardiovascular Systems, Ashland, MA . The Sarns weighted sucker is indicated for stationary use to remove excess fluid from the surgical field.
| Date | Apr 15, 2011 (initiated Feb 25, 2011) |
| Status | Terminated |
| Reason | The firm has received two reports of the Terumo Weighted Flexible Sucker becoming ensnared in cardiac tissue after it was inserted into a ventricle through a valve. The firm has determined that the products's instructions for use are inadequate; the product should not be used to drain a cardiac chamber or inserted through a valve. However this is not stated in the IFU. The firm is conducting |
| Root cause | Device Design |
| Action | SEE UPDATE BELOW IN ALL CAPS. The firm, Terumo Cardiovascular Systems (Terumo CVS), sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 25, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were informed that the product's Indications For Use (IFU) are inadequate; the product should not be used to drain a cardiac chamber or inserted through a valve. However, this is not stated in the IFU. In addition, the customers were instructed to read and understand the changes and clarifications to the product's IFU; review the Medical Devi |
| Quantity | 11,690 |
| Distribution | Worldwide distribution: USA (Nationwide) and countries including: BRAZIL, BELGIUM, BRAZIL, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, and MALAYSIA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.