Recall Z-2568-2011 ยท MUJ System, Planning, Radiation Therapy Treatment

Computerized Medical Systems Inc

XiO Radiation Treatment Planning System, XiO Release 4.3.0 through 4.50.00. Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.

DateJun 15, 2011 (initiated May 13, 2009)
StatusTerminated
ReasonIn XiO, when the beams are renumbered, it is possible to get into a state where the dose displayed for a beam does not match the beam placement on the patient. This can occur when dose is calculated, beams are renumbered, and the dose is re-summed only.
Root causeSoftware design
ActionComputerized Medical Systems, Inc. sent an CUSTOMER ADVISORY letter dated May 2009, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Computerized Medical Systems, Inc have created "patches" to resolve the problem. A User Notice was sent in March 2011, to all users still using the affected software notifying them of the updated software available to correct the problem. This letter was sent by US Mail but no return receipt was included. For further questions, please call (408) 380-8023.
Quantity1116
DistributionWorldwide Distribution -- (USA) nationwide Distribution, including Puerto Rico, and countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Belgium, Bolivia, Bosnia, Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092132XIO RTP SYSTEM
Computerized Medical Systems, Inc., Sep 24, 2009