ClearanceBrief
Recall Z-2175-2011 ยท NSX

Philips Healthcare Inc.

Philips IntelliVue Clinical Information Portfolio software as follows: 865047 IntelliVue Clinical Information Portfolio Release D.03; 865209 IntelliVue Clinical Information Portfolio Release E.0 and E.01; 865300 IntelliVue Clinical Information Portfolio NICU Link

DateMay 10, 2011 (initiated Mar 16, 2011)
StatusTerminated
ReasonEdits to an order can result in unintended scheduled interventions and a patient may receive additional interventions or medication doses that are not intended.
Root causeSoftware design
ActionPhilips Healthcare notified accounts on 3/16/11 via the Urgent Medical Device Correction Notification. Customer notifications in the United States will be delivered via UPS. Customer notifications outside the US will be managed by Philips representatives in each affected geography. Customers will be asked to follow the guidelines in the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notice until they receive their software correction. The correction will be in the form of a software upgrade and will be performed by Philips service personnel. All cust
Quantity387 units
DistributionWorldwide distribution: USA, Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Kenya, Latvia, Malaysia, Netherlands, Norway, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, and Unite
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100272INTELLIVUE CLINICAL INFORMATION PORTFOLIO
Philips Medical Systems, Apr 14, 2010
K992636HP CAREVUE 9000 CLINICAL INFORMATION SYSTEM, MODEL M2331A
Hewlett-Packard Co., Aug 31, 1999