Recall Z-1945-2011 ยท CEC Radioimmunoassay, Free Thyroxine

Beckman Coulter Inc.

Access Free T4 Calibrator, Part Number: 33885, Lot Numbers: 014769 The Access Free T4 Calibrators are intended to calibrate the Access Free T4 assay for the quantitative determination of free thyroxine levels in human serum and plasma (heparin) using the Access Immunoassay Systems

DateApr 11, 2011 (initiated Nov 11, 2010)
StatusTerminated
ReasonThe recall was initiated because the stability claim for Access Free T4 calibrator lot 014769 is not met. The affected T4 calibrator lot can produce quality control and patient results that are erroneously low (by greater than 2 standard deviations). An erroneously low patient result may lead to misdiagnosis and/or delayed treatment.
Root causeOther
ActionThe firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION" (PCA) letter dated November 29, 2010 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: ACTION/RESOLUTION: " Discontinue use of Access Free T4 calibrator lot 014769. " Review historical quality control performance and determine whether any reported patient results were obtained while using a faulty calibration. If results were affected, you will see a shift downward in control recovery that produces out-of-range low QC results. " If pati
Quantity459 units
DistributionWorldwide distribution: USA and countries including: Argentina, Belgium, Bosnia and Herzegovina, Canada, China, Colombia, Czech Republic, French Polynesia, Hong Kong, Ireland, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Oman, Panama, Philippines, Russian Federation, South Africa, Switzerland, T
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982250ACCESS FREE T4 REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER
Beckman Instruments, Inc., Jul 14, 1998