Recall Z-2196-2011 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Edwards Lifesciences Llc

Edwards Aortic Perfusion Cannula with Duraflow Coating, REF: DARH201190TA, 20 Fr. x 28 cm, Sterile EO, Edwards Lifesciences LLC, Irvine, CA 92614. The cannula perfuses blood to the ascending aorta during short-term (<6 hours) cardiopulmonary bypass procedures. The device is a Duraflo coated cannula intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.

DateMay 13, 2011 (initiated Jan 21, 2011)
StatusTerminated
ReasonBonding of the tip to the aortic cannula tubing may decline over time and may contribute to a drop in bond strength. Possible tip separation.
Root causeOther
ActionThe firm, Edwards Lifesciences, sent an "URGENT PRODUCT RECALL - ACTION REQUIRED" letter dated January 21, 2011 it customers via Fed-Ex. The letter described the product, problem and actions to be taken. The customers were instructed to check their inventory for any of the lot numbers; return all unused product from their stock to Edwards; contact their Customer Service Organization at (800) 424-3278 to obtain a Returned Goods Authorization number and replacement product; complete and return the Aortic Perfusion Cannula Recall Response form via fax to: Edwards at (949) 250-3489, and return u
Quantity20
DistributionNationwide distribution: CA only.
FDA recordOpen on fda.gov