Recall Z-1748-2011 ยท NEZ System, Imaging, Gastrointestinal, Wireless, Capsule

Given Imaging Ltd.

Given Imaging Limited, PillCam Express Delivery Device, For use with PillCam SB, endoscopic imaging device. Part numbers FGS-0367 (3-Pak) or FGS-0366 (1-Pak).

DateMar 23, 2011 (initiated Mar 8, 2011)
StatusTerminated
ReasonThe capsule holder can break away from the catheter during endoscopic procedures, necessitating retrieval.
Root causeDevice Design
ActionGiven Imaging, Inc. consignees were notified via FedEx letter on/about 03/08/2011. They instructed to return all affected product. A Customer Response form with return instructions was included to be completed and returned to Given Imaging.
Quantity389
DistributionU.S. Nationwide, Canada, Germany Israel and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101200PILLCAM EXPRESS VIDEO CAPSULES DELIVERY DEVICE
Given Imaging , Ltd., Sep 30, 2010