ClearanceBrief
Recall Z-1874-2011 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Toshiba American Medical Systems Inc

Toshiba Infinix-i, DFP-8000D The expected usage of these devices is for fluoroscopy studies for various anatomical regions of the human body.

DateApr 15, 2011 (initiated Dec 21, 2010)
StatusTerminated
ReasonUnder some conditions during high doses procedures the cumulative air kerma display may reach its maximum capability and fail to display the accrued value.
Root causeRadiation Control for Health and Safety Act
ActionAn Important: Electronic Product Radiation Warning notification was sent to all consignees. The letters described the reason for the recall and identified the affected product. The system software will be modified to allow the shift of the decimal to one position to the right allowing for a display of 99999.9 mGy. This will allow for the retention of the resolution until such time as the system exceeds the 9999.99 mGy range. A Toshiba Service person will contact customers when the software is available. Installed systems with the current version of software ( version 4.23) will begin to r
Quantity220 units Nationwide
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.