Recall Z-1959-2011 ยท KPR System, X-Ray, Stationary

Siemens Medical Solutions USA, Inc

Axiom Aristos FX and FX Plus System Intended use: Stationary x-ray system

DateApr 12, 2011 (initiated Feb 17, 2011)
StatusTerminated
ReasonDuring product monitoring, firm discovered a possible hazard to patients when using the detector alpha rotation of all Axiom Aristos FX and FX Plus systems with serial numbers up to and including 1216. When using the detector alpha rotation, the detector may descend unexpectedly during movement or in vertical or table mode. If this occurs in vertical modes, the possibility exists that the detect
Root causeOther
ActionThe firm, SIEMENS, sent a "Safety Advisory Notice" dated February 15, 2011 to the affected customers via Update Instructions XP016/10/S on February 17, 2011. The notice described the product, problem and actions to be taken. The customers were instructed to immediately instruct their personnel accordingly; ensure that the safety advisory is placed in the systems's instructions for use; if they have sold this device/equipment and it is no longer in their possession, forward this safety notice to the new owner of the device/equipment; inform SIEMENS about the new owner of the device/equipment,
Quantity101
DistributionNationwide Distribution: USA including states of: AZ, AK, CA, CO, FL, GA. IL, IN, IA, KS, KY, ME, MI, MN, MS, MO, NE, NV, NJ, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013826AXIOM ARISTOS FX (MULTI PURPOSE RADIOGRAPHY SYSTEM)
Siemens Medical Systems, Inc., Feb 1, 2002
K061054AXIOM ARISTOS FX PLUS
Siemens Medical Solutions USA, Inc., May 9, 2006