Recall Z-1942-2011 ยท LLZ System, Image Processing, Radiological

GE OEC Medical Systems, Inc

Receiver Sensor Cables, Part Numbers: 1001989, 1001990, 1004069, 1007907-NAV, GE Healthcare Surgery, Salt Lake City, UT 84116. Cables are used with InstraTrak Navigation Systems, Models IT3000, IT2500, IT2500+, IT3500, IT3500+. InstaTrak System is intended as an aid to the surgeon for precisely locating anatomical structures anywhere on the human body during either open or percutaneous procedures.

DateApr 11, 2011 (initiated Feb 24, 2011)
StatusTerminated
ReasonSensor cables may suffer material degradation when exposed to certain sterilization procedures.
Root causeComponent design/selection
ActionGE OEC Medical Systems, Inc sent an Urgent Recall Notice dated March 1, 2011, by certified letter on February 24, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised if their Transmitter Sensor Cables or Receiver Sensor Cables have been exposed to any STERRAD sterilizer other than STERRAD 1005, discontinue use of these sterilizers for this product. The Transmitter Sensor Cables should be inspected for signs of material degradation. Further information is available at 800-874-7378.
Quantity708 cables
DistributionWorldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI and WV and the countries of Austria, Canada, China, Denmark, England, France, Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040050INSTATRAK SYSTEM WITH MULTIPLE DATASET NAVIGATION, MODEL IT3500
Ge Medical Systems Navigation and Visualization IN, Mar 24, 2004