Recall Z-1941-2011 ยท LLZ System, Image Processing, Radiological
GE OEC Medical Systems, Inc
Transmitter Sensor Cables, Part Numbers: 1002008, 1004587, 1007914-NAV, GE Healthcare Surgery, Salt Lake City, UT 84116. Cables are used with InstraTrak Navigation Systems, Models IT3000, IT2500, IT2500+, IT3500, IT3500+. InstaTrak System is intended as an aid to the surgeon for precisely locating anatomical structures anywhere on the human body during either open or percutaneous procedures. It is
| Date | Apr 11, 2011 (initiated Feb 24, 2011) |
| Status | Terminated |
| Reason | Sensor cables may suffer material degradation when exposed to certain sterilization procedures. |
| Root cause | Component design/selection |
| Action | GE OEC Medical Systems, Inc sent an Urgent Recall Notice dated March 1, 2011, by certified letter on February 24, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised if their Transmitter Sensor Cables or Receiver Sensor Cables have been exposed to any STERRAD sterilizer other than STERRAD 1005, discontinue use of these sterilizers for this product. The Transmitter Sensor Cables should be inspected for signs of material degradation. Further information is available at 800-874-7378. |
| Quantity | 708 cables |
| Distribution | Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI and WV and the countries of Austria, Canada, China, Denmark, England, France, Germany |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K040050 | INSTATRAK SYSTEM WITH MULTIPLE DATASET NAVIGATION, MODEL IT3500 Ge Medical Systems Navigation and Visualization IN, Mar 24, 2004 |