Recall Z-2021-2011 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

MAGNETOM Avanto, Espree, Trio A Tim System. Nuclear magnetic resonance imaging.

DateApr 18, 2011 (initiated Mar 3, 2011)
StatusTerminated
ReasonFirm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
Root causeOther
ActionSiemens Medical Solutions Inc., issued a Customer Safety Advisory Notice dated November 24, 2010, to affected customers via Update Instructions MR029/10S. Affected customers will be notified via hand delivered letter by Siemens Service Engineers at the time of inspection. A future Update instruction will be released for installation of detectors in appropriate locations on systems to switch off the power electronics in the event of a malfunction in the power supply of the gradient coils. Further questions regarding this recall please call (610) 448-3237.
Quantity1419
DistributionProduct was distributed domestically to hospitals and imaging centers in the District of Columbia, Puerto Rico, AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013586MAGNETOM TRIO SYSTEM
Siemens Medical Systems, Inc., Dec 28, 2001
K032428MAGNETOM AVANTO MR SYSTEM
Siemens Medical Solutions USA, Inc., Oct 16, 2003
K041112MAGNETOM ESPREE SYSTEM
Siemens Medical Solutions USA, Inc., Jul 21, 2004
K050200MAGNETOM TRIO A TIM SYSTEM
Siemens Medical Solutions USA, Inc., Feb 28, 2005