Recall Z-0905-2011 ยท IZO Generator, High-Voltage, X-Ray, Diagnostic

Philips Healthcare Inc.

BuckyDiagnost general purpose x-ray system with Optimus Generator. General Purpose Diagnostic Radiography.

DateMar 4, 2011 (initiated Jan 17, 2011)
StatusTerminated
ReasonAffected systems were determined not to prevent run-on of a selected technique factor if the technique selector and the exposure control were activated simultaneously.
Root causeRadiation Control for Health and Safety Act
ActionA Customer Information BuckyDiagnost Diagnostic X-Ray System letter, dated January 6, 2011, was sent to customers. The letter identifies the affected product, the reason for the recall, actions that are to be taken by customers, and actions that are planned by the firm. Customers are advised to make sure that the desired exposure values are displayed and that they have released the technique button BEFORE they start the exposure by pressing the exposure switch whenever manually changing the exposure parameters and in order to avoid the problem. Philips will issue a Field Change Order (FCO)
Quantity1810 units in the US
DistributionWorldwide Distribution