ClearanceBrief
Recall Z-1843-2011 ยท HQF Laser, Ophthalmic

Ellex Medical Pty Ltd

Ellex Tango LT5106-T ophthalmic laser (not sold in USA)

DateApr 12, 2011 (initiated Feb 15, 2011)
StatusTerminated
ReasonUnintended laser emission while the user was moving the joystick for targeting.
Root causeProcess control
ActionEllex notified all distributors that have received the affected devices per Global Customer Service Bulletin GSB-11 -01a, dated 25 February 2011.
Quantity679 units
Distributionworldwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.