ClearanceBrief
Recall Z-2078-2011 ยท CEM Electrode, Ion Specific, Potassium

Beckman Coulter Inc.

UniCel DxC 800 The UniCel DxC SynchronClinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).

DateApr 26, 2011 (initiated Oct 22, 2010)
StatusTerminated
ReasonThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Root causeEnvironmental control
ActionThe firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated October 20, 2010 with attached PCA Response Form (via US Postal Service for US Customers) to all customers who purchased Synchron LX and UniCel DxC Clinical Systems Operating Software (all versions). The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: -After modifying any UDR wavelength, they MUST reboot to apply the update of the polychrome correction. -If the modified UDR requires calibration, recalibrate after rebooting. - Complete a
Quantity4879 units total (3490 in US)
DistributionWorldwide distribution: USA (nationwide) and countries including:Angola, Australia, Belgium, Bulgaria, China, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Egypt, Eritrea, Finland, France, French Polynesia, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Ku
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042291UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS
Beckman Coulter, Inc., Nov 12, 2004