ClearanceBrief
Recall Z-1712-2011 ยท CAE Airway, Oropharyngeal, Anesthesiology

King Systems Corp.

CRQ424, Cardiac ResQ Kit, Quantity 3 units, KING SYSTEMS, A Consort Medical Company, Noblesville, IN 46060, www.kingsystems.com, 317-776-6823. Advanced Circulatory Systems, Inc., Eden Prairie, MN, www.advancedcirculatory.com, 877-7377763. This Oropharyngeal Airway is intended for airway management by providing a patent airway to allow patient ventilation. The sterile lubricant is used in Assemb

DateMar 17, 2011 (initiated Jan 14, 2011)
StatusTerminated
ReasonKing Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be "inadequately sterilized" and use of this product may result in "patient infection". Because this particular sterile lubricating jelly is packaged in the KING LT-D Kit and KING LTS-D Kit, King Systems is sending a Notice of the Reca
Root causeProcess control
ActionThe firm, King Systems, sent an "URGENT PRODUCT RECALL" letter dated January 14, 2011 to its customers. King Systems sent a follow-up "URGENT PRODUCT RECALL" letter dated March 3, 2011 which "SUPERSEDES" the January 14, 2011 recall letter with attached instructions to its customers. The letters described the product, problem and actions to be taken by the customers. The customers were instructed to notify all their purchasers of this recall; immediately cease sales and use of the affected products; identify all inventory on hand or returned to them by their customers; REMOVE Triad Group lubric
Quantity577,426 total units
DistributionWorldwide distribution: USA including: CA,CO, FL, ID, IN, KS, NC, NH, NY, OH, PA, PR, SC, TN, and TX; and Country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021634KING LT, MODELS # KLT 100, KLT 101, KLT 102, KLT 103, KLT 104, KLT 105
King Systems Corp., Jan 9, 2003
K033186KING LT-D DISPOSABLE OROPHARYNGEAL AIRWAY, MODELS KLT 203, KLT 204 AND KLT 205
King Systems Corp., Apr 14, 2004