Recall Z-1664-2011 ยท DRM Compressor, Cardiac, External

Jolife AB

LUCAS 2 Chest Compression System. The device's Instruction for Use is labeled in part: "LUCAS 2 Chest Compression System***jolife***www.lucas-cpr.com***a product by JOLIFE***MANUFACTURER, MAIN OFFICE JOLIFE AB Scheelevagen 17 SE-223 70 LUND Sweden***www.jolife.com***". Designed to uninterrupted chest compressions at a consistent rate and depth.

DateMar 16, 2011 (initiated Feb 8, 2011)
StatusTerminated
ReasonLUCAS 2 Chest Compression System may have malfunctions resulting from the failure of a hood probe on the device that connects to the battery. The device may stop in fail safe mode and the LEDs will start flickering or the device will switch off.
Root causeOther
ActionMedical Device Correction letter, dated February 2011, was sent to customers and states that Jolife AB with the servicing support assistant of Physio-Control is conducting a voluntary upgrade of the LUCAS 2 Chest Compression System with the Serial number range of 30090016 through 30101932. Consignees were advised to keep LUCAS 2 in service and continue to use it in accordance with the Instruction for Use, Section 3.8-Safety Precautions. WARNING-MALFUNCTION. If there are interruptions, or the compressions are not sufficient or something unusual occurs during operation: Push ON/OFF for 1 secon
Quantity1843 units
DistributionWorldwide Distribution -- United States, Australia, Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Saudi Arabia, Singa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090422LUCAS 2
Jolife AB, Jun 17, 2009