Recall Z-1589-2011 ยท HRX Arthroscope

Biomet, Inc.

Biomet INNERVUE DISPOSABLE SCOPE KIT 1.2MM WITH CANNULA SET, PRODUCT NUMBER 922141, STERILE

DateMar 7, 2011 (initiated Feb 10, 2011)
StatusTerminated
ReasonThe cannula could not be connected to the scope and scopes disengage during use. The outer hub of Innervue scopes may become detached from the inner hub of the scopes during assembly of the scope to the Innervue hand piece. This condition may prevent the Innervue scope from being securely attached to the Innervue hand piece.
Root causeProcess design
ActionOn 2/10/2011, Biomet issued an "URGENT MEDICAL DEVICE NOTICE" to their Direct Accounts instructing customers to identify and discontinue use of the recalled products and return them to Biomet. Response forms were provided.
Quantity264
DistributionDistribution USA nationwide and to the Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072879INNERVUE DIAGNOSTIC SCOPE SYSTEM