ClearanceBrief
Recall Z-2035-2011 ยท HTT Burr, Orthopedic

The Anspach Effort, Inc.

***REF MCA5-15SB ***1.5 mm Fluted Ball, 5.6 cm. Use with Micro Curved Attachment (MCA).***ANSPACH***Sterile, Sterilized using Irradiation***Rx Only*** Made in the USA***Manufacturer: The Anspach Effort, Inc. 4500 Riverside Drive, Palm Beach Gardens, FL 33410 USA Tel: (800) 327-6887/ +1 561 627 1080***Anspach Europe Ltd. Aston Court, Kingsmead Business Park Frederick Place High Wycombe, HP11 1LA

DateApr 20, 2011 (initiated Sep 29, 2010)
StatusTerminated
ReasonAnspach Effort, Inc. Palm Beach Gardens, FL is recalling their Sterile Bone Cutting Burs that was manufactured with incomplete flutes which could cause degraded cutting performance and in some cases not cut at all. Anspach has not received any complaints related to this problem.
Root causeOther
ActionThe firm, Anspach, sent an "URGENT: Medical Device Product Removal" letter dated October 6, 2010, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Screen their inventory immediately for any of the part numbers listed in the letter. Do not include any product received after October 7, 2010. 2) If they have any of the products listed, contact Anspach Customer Support at (800) 327-6887 to arrange for their immediate return and replacement. 3) Complete and return the attached Customer Reply Form indicating h
Quantity271 cutting burs affected)
DistributionWorldwide distribution: USA including states of: AZ, CA, FL, GA, IL, IN, LA, MA, and MI; and countries of: Australia, Belgium, England, France, Italy, Japan, South Africa, Spain and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.