Recall Z-1819-2011 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Extended Brilliance Workstation (EBW) a component of the Computed Tomography X-Ray System Model Numbers EBW (CT) 728260 & EBW (Portal) 728269. The Extended Brilliance Workplace (EBW) is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This d

DateMar 24, 2011 (initiated Oct 3, 2009)
StatusTerminated
ReasonDuring an FDA inspection on 10/1/2010 the firm was notified by FDA that their field action they initiated in Oct 2009, constituted a recall. The firm determined that the functional analysis feature in the Comprehensive Cardiac Analysis (CCA) was displaying incorrect values on the polar fields. The calculated results for the myocardial wall thickening was incorrect in that the numbers displayed f
Root causeOther
ActionPhilips sent an URGENT - Field Safety Notice letter dated September 29, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to be aware that the numbers displayed for "Wall Thickening" actually represent a reduction in wall thickness from ES to ED. Philips will be providing a sotware update to resolve the issue. For further information or support customers were instructed to contact their local Philips representative: Customer Care Service Center in the U.S. at 1-800-722-9377, option 5: Diag
Quantity2,068 units
DistributionWorldwide Distribution - USA including the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX , UT, VA, VT, WA, WI, WV & WY and the following countries: ARGENTINA, AUSTRALIA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001